Synergy Nano Technical Manual, Revision -
Page 134
Example 2
At low atmospheric pressures, the heaters in most altitude chambers are turned off so they don’t overheat
when convection cooling capacity is reduced. In this example we will create an alarm that senses Hi Res
Input 3 (Torr) and opens Relay 2 when the value is less than 30 Torr. This alarm is named “Heater Safety
Shutoff” and displays “Heater Safety Shutoff” when the threshold is reached.
Open the User Alarm folder and press the
Add Alarm button on the Setup screen shown
at left.
Select the Olympic Module and Analog 1, the
Hi Resolution Analog input.
This screen shows the Sensor code for the
Olympic Module and Analog input 1.
Since we want the alarm to trigger when the
pressure is less than 30 Torr, we select the
Page view 133