17.8 Controller Setup
Once the Synergy Nano and output boards are mounted and wired check the tightness of all connections.
Then apply power to the system and configure the controller. To configure the controller:
1. Startup the controller and select the chamber type from the Setup/Chamber setup folder.
2. Check the Input calibration for all inputs and the sensor selection for each channel. Change as
necessary.
3. Setup the channel alarms and any other alarms (see the Section 3.0 - Safety
)
4. Setup Logging; Log interval, log data, log enable. (See the Section 6.11 - Logging
)
5. Enter the WebTouch Remote, Cascade Control and Pressure Control Registration keys as
necessary.
17.9 Alarm System Test
Test all alarm systems. Verify that test chamber shut down occurs and power is removed from heaters,
compressor, etc. when any sensor is disconnected. Verify that test chamber shut down occurs when the
secondary limit controller alarm limits are reached. (See the Section 3.0 - Safety
)
17.10 Control System Verification
Verify control system performance for transient and steady state control. Tune the PID settings as
necessary. (See Section 6.3 – PID) Consult the factory for additional assistance with chamber tuning.