Synergy Nano Technical Manual, Revision -
Page 37
Channel Calibration
Adjust the Channel Offset (b) and Gain
%(m) to accommodate channel specific
sensor errors such as those caused by the
sensor placement. General sensor
calibration should be setup using Input
Calibration screens shown in the next
section.
To adjust the Channel Calibration
parameters, select the parameter, press
Change and enter the new value in the
number pad.
Channel Absolute Alarms
The Synergy Nano provides High and Low
limit Alarms.
To adjust the absolute Low Alarm limit,
select Low Alarm, Channel n, press
Change and enter the new value in the
number pad. (n is the channel)
To adjust the absolute High Alarm limit,
select High Alarm, Channel n, press
Change and enter the new value in the
number pad. (n is the channel)
The Absolute Alarm limits can be disabled
automatically when the Channel is off
using the Ignore Alarm When Off
parameter.
Channel Deviation Alarms
The Synergy Nano offers Deviation Alarm
Limits.
To adjust the Deviation Low Alarm limit,
select Deviation Low Alarm, press Change
and enter the new value in the number
pad.
To adjust the Deviation High Alarm limit,
select Deviation High Alarm, press
Change and enter the new value in the
number pad.
The Deviation Alarm limits can be disabled
or delayed using the Deviation Alarms
Enabled and Deviation Alarm Delay
parameters.