Synergy Nano Technical Manual, Revision -
Page 80
Switching Module Configuration
The Olympic board drives all of the outputs for the chamber thru
solid state switches called Switching Modules (SM). In some
cases there is more than one way to connect a specific output.
This provides flexibility when wiring the chamber to support new
and retrofit installations. The figure at the right shows the
different the ways Olympic board can be connected to the
various SM boards.
For example, note that the 3SM-Event outputs can be connected
two ways; directly to the Olympic board P6 connector or thru the
1SM- J5 connector as shown at the right
In addition, some chamber definitions provide the same function
to more than one output. This is referred to as mirroring.
For example, in the Generic Temperature Humidity configuration
the “Humidify” output is available on 2SM-2 and SSR-2. The
SSR outputs emulate the VersaTenn III SSR outputs which
simplifies VersaTenn controller retrofits. To further support
retrofits, the 5-Channel output board is wired to emulate the
wiring of the VersaTenn III SSR outputs.
Installation section 17.0
describes the SSR outputs for retrofit
configurations in greater detail.
The Screenshot below shows the position of each switching
module on the EVENTS/Digital Output Screen.
1SM
1 Thru 8
Alarm
Relays
1 and 2
SSR
Out
p
uts
2SM
1 Thru 6
3SM
1 Thru 6
1SM
9 Thru 10
1SM
11 Thru 12