Synergy Nano Technical Manual, Revision -
Page 300
Description Command Root Command
Usages
Command Syntax Range,
Units
Command
Example
Response Example
CAL1 Query ? CAL1 -50 to 50 C -90 to 90 F ? CAL1 10.00
Low Alarm,
Ch 1
A1L A1L Set = A1L ARG1 -200 to 500 C -326 to 932 F = A1L -200 OK
A1L Query ? A1L -200 to 500 C -326 to 932 F ? A1L -200
High Alarm,
Ch 1
A1H A1H Set = A1H ARG1 -200 to 500 C -326 to 932 F = A1H 500 OK
A1H Query ? A1H -200 to 500 C -326 to 932 F ? A1H 500
Ch 1 Alarm
Enabled
While Off
IGNORE_CH1_AL
M
IGNORE_CH1_
ALM Set
= IGNORE_CH1_ALM
ARG1
ARG1:
0 - Outputs Off
1 - Outputs On
= IGNORE_CH1_ALM 1
OK
IGNORE_CH1_
ALM Query
? IGNORE_CH1_ALM
ARG1
0 - Outputs Off
1 - Outputs On
? IGNORE_CH1_ALM
1
Ch 1 Gain
GAIN1 GAIN1 Set = GAIN1 ARG1 82 - 120%
= GAIN1 90 OK
GAIN1 Query ? GAIN1 83 - 120%
? GAIN1 90.00
Ch 1
Deviation
High Alarm
A1DH A1DH Set = A1DH ARG1
ARG1:
0 - 500C
= A1DH 20 OK
A1DH Query ? A1DH 0 - 500C
? A1DH 20.0
Ch 1
Deviation
Low Alarm
A1DL A1DL Set = A1DL ARG1
ARG1:
0 - 500C
= A1DL 10 OK
A1DL Query ? A1DL 0 - 500C
? A1DL 10.0
Ch 1
Deviation
Alarms
Enabled A1DE A1DE Set = A1DE ARG1
ARG1:
0 - Disabled
1 - Enabled
= A1DE 1 OK
A1DE Query ? A1DE
0 - Disabled
1 - Enabled
? A1DE 1
Ch 1
Deviation
Alarm Delay
A1DT A1DT Set = A1DT ARG1
ARG1:
0 - 7200 Seconds
= A1DT 3600 OK
A1DT Query ? A1DT 0 - 7200 Seconds
? A1DT 3600
Ch2 Sensor
Select
CH2SENSOR CH2Sensor Set = CH2SENSOR ARG1 ARG1 - ID of the Sensor. 100 - 999. = CH2SENSOR 130 OK
CH2 Sensor
Query
? CH2SENSOR See the user manual for numeric
codes.
? CH2SENSOR 130
Calibration
Ch 2
CAL2 CAL2 Set = CAL2 ARG1 -50 to 50 C -90 to 90 F
-50 to 50 %RH
= CAL2 10.0 OK
CAL2 Query ? CAL2 -50 to 50 C -90 to 90 F
-50 to 50 %RH
? CAL2 10.00