7.6 Clean Dat Utility
The Clean Dat utility removes all the chamber data (dat) files from your Synergy Controller. After the Clean
Dat program runs and the system reset, all of the controller settings are reset to their default values. You
may want to use the Clean Dat Utility if your controller won’t boot. Deleting the dat files may resolve the
issue.
Before you run the Clean Dat utility make sure to record all of the settings (PID, Special Functions, LValues,
etc) for the controller. See the Settings List in Section 6.6 within the Setup chapter of this manual for a
helpful table that you can use to record your chamber settings. Alternatively you can use Tidal Engineering’s
SimpleComm communications application to record the settings via Ethernet or Serial Communications and
restore them after you run the Clean Dat utility. The SimpleComm application is discussed elsewhere in this
manual and is available on the Resource CDROM and on Tidal Engineering’s website. Running the Clean
Dat utility will NOT delete your chamber’s profile program files (files you created using the Program Screen).
To clean your dat files:
1. Insert the Synergy Controller Clean disk into the USB drive on your chamber.
2. Either reset the controller or cycle the power on the chamber.
3. The Clean Dat will run when the controller reboots and will delete the data files.
4. When complete, and the USB light stays off, remove the USB from the drive.